Weak Suction Troubleshooting

Suction suddenly dropped or gradually weakening? Follow these five steps to troubleshoot — most issues can be resolved without professional help.

Troubleshooting Steps (Follow This Order)

1

Check the Hydrophobic Filter

This is the most common cause of suction loss. When the hydrophobic filter gets wet, it blocks airflow, causing suction to stop or drop significantly. Drive DeVilbiss explicitly states: "A wet hydrophobic filter will stop airflow." Remove the filter and check if it is wet or discoloured. If so, replace it immediately with a new one. Do not attempt to blow-dry or wash and reuse.

Solution: Replace with a new hydrophobic filter → Filter Accessories

2

Check the Tubing

The ATMOS manual lists "kinked tubing" as a common cause of suction loss. Inspect the entire tubing for: kinks, cracks, contamination or air leaks. Aged tubing becomes hard and brittle, cracking more easily.

Solution: Replace with new tubing → Tubing Accessories

3

Check the Collection Jar

An overfull jar reduces suction. Check if the jar is near capacity. Also inspect the lid gasket — a poor seal causes air leaks.

Solution: Empty the jar and check the gasket seal

4

Check the Connections

Are all tubing-to-machine connections secure and sealed? Is the connector loose or damaged? An incorrectly sized connector causes air leaks.

Solution: Reconnect and confirm seal; replace connector if needed → Connector Accessories

5

Professional Inspection

If suction is still weak after the above steps, the pump may be worn or there may be an internal mechanical issue. A qualified technician should inspect the unit.

Solution: Book a suction pressure test or repair service

Suction Specifications by Model

If your machine cannot reach these specs, performance is below standard:

ModelStandard Max VacuumStandard Airflow
ATMOS LC27-73 kPa27 L/min
ASKIR 230-75 kPa16 L/min
ASKIR 36 BR-80 kPa36 L/min
VACUAIDE 7325P-550 mmHg27 LPM

FAQ

Replaced the filter but still no suction?

Check if you replaced both the bacterial filter and the hydrophobic filter. Some models have two separate filters. Also inspect tubing and connections.

Suction weaker on battery than on mains?

Low battery charge can reduce suction. Fully charge first, then test. If still weaker at full charge, the battery may be degraded and need replacement.

Brand new machine has weak suction?

Weak suction on a new machine is usually an assembly issue: filter not installed correctly, tubing not properly connected, or jar not sealed. Reassemble following the manual.